Blood microsampling method quantified nine cannabinoids and metabolites in specimens

Paper title: Non-psychotropic phytocannabinoids, Δ9-THC and their metabolites by means of volumetric dried blood microsampling

Journal of Cannabis Research Authentic whole-blood specimens; number and donor details not stated

The short answer

Researchers developed and validated a blood microsampling method that quantified nine cannabinoids and metabolites, with room-temperature stability confirmed for at least 90 days. It was tested on authentic blood specimens, but the abstract does not report participant numbers or clinical outcomes; it establishes an analytical method, not a health benefit.

Design
Lab analysis (chemistry)
People
Authentic whole-blood specimens; number and donor details not stated
Evidence
Lab and cells
Cells or chemistry, not people
Source
Peer reviewed
Journal of Cannabis Research

What it can't tell you

The abstract does not state limitations.

What the study found

  1. A validated HPLC–MS/MS method simultaneously measures nine cannabinoids and metabolites in volumetric dried blood microsamples.
  2. The method showed at least 90 days of room-temperature stability for the complete analyte panel.
  3. The method successfully quantified the full panel in authentic blood specimens, supporting its potential use in monitoring and research.

Each finding is supported by the abstract.

The full summary

This study developed and fully validated an HPLC–MS/MS method to measure nine cannabinoids and metabolites in whole blood collected with a fingerstick-based volumetric dried blood device. The panel included CBD, Δ9-THC, CBN and other cannabinoids and metabolites; the method used separate extraction procedures to accommodate different expected blood concentrations.

The approach was validated for measures including accuracy, precision, stability and independence from hematocrit. The abstract reports room-temperature stability for at least 90 days and says the method successfully quantified the full panel in authentic blood specimens. This is an analytical-method study, not a treatment trial: it supports potential use in forensic, toxicological and therapeutic drug monitoring, but does not establish clinical benefits or outcomes.

Original abstract

The growing therapeutic use of cannabinoid-based formulations and the expanding heterogeneity of available cannabis products have created a pressing need for analytical methods capable of simultaneously monitoring a wide spectrum of pharmacologically relevant cannabinoids in blood. Currently, no fully validated HPLC–MS/MS method is available for the comprehensive analysis of a multi-cannabinoid panel in whole blood collected via volumetric dried microsampling. To address this gap, we report the development, optimisation and full validation of an HPLC–MS/MS method for the simultaneous quantification of nine cannabinoids, namely cannabidiol (CBD), 7-hydroxy-CBD (7-OH-CBD), 7-carboxy-CBD (7-COOH-CBD), 6α-hydroxy-CBD (6α-OH-CBD), cannabinol (CBN), cannabigerol (CBG), Δ9-tetrahydrocannabinol (Δ9-THC), 11-hydroxy-THC (11-OH-THC) and 11-nor-9-carboxy-THC (THC-COOH), in whole blood collected as advanced volumetric dried blood microsamples using the hemaPEN® device. Chromatographic separation was achieved on a C8 column with electrospray ionisation in polarity switching. A dual-protocol extraction strategy was introduced to address the wide dynamic range of cannabinoid concentrations in blood: analytes with high expected concentrations (THC-COOH, CBD, 7-OH-CBD and 7-COOH-CBD) were quantified from a single hemaPEN disc (2.74 µL; LLOQs 2.0–5.0 ng/mL), while Δ9-THC, 11-OH-THC, 6α-OH-CBD, CBN and CBG, requiring enhanced sensitivity, were extracted from all four pooled discs (10.96 µL; LLOQ 0.5 ng/mL for Δ9-THC, 11-OH-THC, 6α-OH-CBD, CBN and CBG). This strategy uniquely exploits the four-replicate architecture of the technology, which generates four volumetrically identical dried blood spots by capillary action from a single fingerstick. The method was validated for selectivity, linearity, precision, accuracy, extraction yield, matrix effect, hematocrit independence and stability. Extraction yield ranged from 67.8% to 81.7%; IS-normalised matrix effect ranged from 89.6% to 98.6% across six donor matrices. Room-temperature stability was confirmed for at least 90 days for the complete analyte panel. The method was applied to authentic blood specimens, demonstrating successful quantification across the full nine-cannabinoid panel. The present method provides a validated, logistically practical tool for forensic and toxicological cannabinoid monitoring, therapeutic drug monitoring (TDM) in patients receiving cannabinoid-based treatments, and pharmacokinetic characterisation in clinical and preclinical research.

Paper details 7 authors, keywords and identifiers

Authors

  • Sobia Noreen
  • Roberto Mandrioli
  • Roberta Di Lecce
  • Chiara Pia Iattoni
  • Mattea Carmen Castrovilli
  • Michele Protti
  • Laura Mercolini

Subjects

Identifiers

PMID
Not in PubMed yet
DOI
10.1186/s42238-026-00503-y

Your stash, decoded.